What should I check if there is no output from the high-frequency electric knife?

Source: 本站 | Date of issue: 2021-08-30 13:18:32 | Views: 4775

Abstract:

1. Does the power socket (or extension strip) have electricity?2. Is the main power cable disconnected?3. Is the power fuse of the electric knife blown due to long-term oxidation?4. Is the power switch in place?5. Has the alarm of the one-time use…

1. Does the power socket (or extension strip) have electricity?

2. Is the main power cable disconnected?

3. Is the power fuse of the electric knife blown due to long-term oxidation?

4. Is the power switch in place?

5. Has the alarm of the one-time use of neutral* (electrosurgical negative* board) been eliminated? Have you not confirmed the contact quality of the double * disposable neutral electric * (negative * plate of the electrosurgical knife)?

6. Is the start switch (manual control, foot pedal) connected to the machine, is it damaged or the lead wire is broken?

7. Is there any disconnection in the cables of the disposable electrosurgical pen, foot-controlled electrosurgical pen, or double * electrocoagulation tweezers?

8. Is the foot switch inserted in the wrong position? There are two sets of foot switch sockets on the back of high-end models. Check whether the surgical accessories correspond to them. Is the mode selected correctly?

9. If there is no output after eliminating the above, it is a host failure, you should contact the host manufacturer's after-sales service.